SCS-5050

Atmospheric Consistometer of The Cement Mortars
SCS-5050
Atmospheric Consistometer Of The Cement Mortars
The atmospheric consistometer consists of rotating the container with cement mortar with a given angular velocity and a vane device occurs torque, proportional to the viscosity of the solution. The spring of the measuring system (potentiometer) is twisted by the angle proportional to the torque and the viscosity of the solution, and the spring is kinematically connected with the axis of the bladed device. The angle of rotation of the spring is interpreted to measure the consistency of cement mortar with the help of the measuring system. The container of the consistometer cement solution prepared in accordance with ISO 10426-1: 2000 to measure thickening time. The container is placed in a bath of the device and the billed device must be in cement mortar.
The container begins rotate. Then turn on the heating device. In the process of testing regulate the temperature in the container with cement mortar in accordance with specified test mode. 
Time from start of mixing the cement resonates as the thickening time of the cement slurry to the consistency of 100 BC.
 
SCS-5050 Technical Specification:
  • determine the range of consistencies from 0 to 100 (Bc);
  • the limits of permissible reduced error of temperature determination are ± 0.1 ° C of the upper limit of determination;
  • the frequency of rotation of the beaker cement mortar 150 rpm;
  • the maximum heating temperature cement mortar is 95 ° C;
  • heater power is 1800 watts;
  • control of the heating of the sample of cement solution is carried out a temperature controller on the microprocessor, the results of which are shows on the touchscreen display;
  • working fluid recommended by the manufacturer is mineral oil ShelI-Ondina927 (contributes to long operation life of the actuator part);
  • the working fluid is an antifreeze for heating systems with a boiling  point ≥ 105 ° C, which does not contain ethylene glycol which as the main component is used glycerin. Antifreeze must meet the requirements of GOST 28084-89 (the use of this working fluid period of operation the operating parts of the device is reduced). - pressure and flow of a fluid cooling 0,15-0,6 MPWP / 2l/min;
  • the type of current is alternating and single-phase;
  • the voltage is 220V (-15 + 10)%;
  • the frequency is 45-65Hz;
  • the grounding.
  • atmospheric consistometer model 5050 climatic design UHL 4.2 (NF);
  • the enviroment temperature from -10° to 35 ° C;
  • relative humidity is up to 65% at 20 ° C.