TEM-100

TEM-100

TEM-100
TEM-100 compact transmission electron microscope for the widespread use of biomedical research, research in nano science and technology fields, as well as for the training of higher education specialists.
 

Management system
      The microscope is controlled by a computer. Software in Windows operating environment allows you to manage the device and display an image while operating the CCD-camera. Controls allow work on the device with minimal distraction of the operator on the monitor screen. Boards can move that allows the operator more convenient to place them on the table.
   


Research in medicine: rat muscle tissue

Photographic system
• vacuum camera with recording on a flat film measuring 90×120 mm or 80×100 mm;
• automatic photographic exposure meters;
• on the film indicated the experiment number and an increase in the length or the camera (for more      information about the image is stored in the database of the computer);
• the ability to install the lower image display system on the screen with the CCD-camera
 

 

 

       
                                                 
                                                Advantages TEM-100:
  • high quality electron microscopic images;
  • easy control the device using the computer;
  • existense high contrast mode for biomedical abjects;
  • original design vacuum camera;
  • easy and reliable vacuum system ;
  • output electron microscopic images on the screen using CCD-camera and image editing with
  • special programs.

                                               Technical Specification:


Parameters
 
Basic Mode
 
Contrast Mode
 
Resolution on points, nm
 
0,4
 
0,5
 
The range of the electron-optical
increases
 
50...600000
50...430000
The range of the diffraction length
chamber, mm
200...2000

280...2800
 
The accelerating voltage, kV
25...100
with the adjustment steps, kV
 
0,05; 1; 5
Power supply (1 phase), V
 
220
Frequency, Hz
 
50/60
 

   The microscope may operate in the following modes: low magnification; large increases in the ground or contrasting modes of operation; microdiffraction mainly or contrasting modes.
The microscope is designed for indoor or partially conditioned air conditioned at a temperature of (293 ± 278) ° K and relative humidity (65 ± 15)% at a temperature of (293 ± 278) ° K . Power supply - single phase is 220 V, frequency 50 Hz